EXCELLENT QC DESIGN (Mean σ = 5.58). Troponin I on Abbott Alinity i: Simple QC rules with 2 control levels run daily are sufficient. Recommended: 1₂.₅s / 2₂s / R₄s / 4₁s. High sigma processes rarely produce unreported errors with standard QC.
References:
Westgard JO. Six Sigma Quality Design & Control, 2nd ed. 2006 ·
Westgard JO, et al. Clin Chem 1981;27:493–501 ·
CLSI C24-A4 — Statistical QC for Quantitative Measurement Procedures Formula: σ = (TEa − |Bias|) / CV ·
Recommendation based on σ tier: ≥6=World Class, 5–6=Excellent, 4–5=Good, 3–4=Marginal, <3=Poor Note: QC rule recommendations are based on probability theory.
Final QC design should be reviewed by the laboratory director per CLIA §493.1256 and CAP GEN.55500.